Unboxing
TRIAL INFORMATION
If you are undertaking a trial, you will need to retain the original box you received your product in, the inner packaging and correctly re-box your product.
Ensure that all components of the product have been thoroughly cleaned and free of any contaminants before returning at the end of your trial period.
How to Use
Assemble the Face Pad Support onto the Adjustable Assembly
Remove the thumbscrew from the Adjustable Assembly, take the Face Pad Support and assemble them together placing the conical washer first followed by the thumbscrew.
Place the soft blue pad on top of the plastic face support and push until the 4 tips are inserted in the corresponding cavities. Make sure all components are safely assembled.
Set up the Adjustable Assembly
Take the assembled Face Pad Support and Adjustable Assembly and align with the Base Board recess. Tilt slightly inward before pushing down into the recess of the Base Board. Straighten it up and push downward until it clips into place.
Set up the Base Board
The Prone Support is compatible with standard rectangular tables.
Ensure the patient is on the table and in prone position prior to product placement. Lift the mattress and slide the Base Board under until the protrusion on the underside contacts the end of the table and centralise.
Height Adjustment
Turn the Locking Lever 90° to the right or left direction to unlock the Adjustable Assembly. Adjust to maximum or minimum height and lock it by rotating the Locking Lever 90° down to a vertical position.
Adjusting the angle of the Face Pad Support
Slightly release / untighten the Thumbscrew located on top of the Adjustable Assembly under the Face Pad Support and adjust the angle until the patient’s head is in a comfortable position.
Positioning the Patient
Place the patient’s arms on the arm supporting surface according to their level of mobility being careful not
to staring the shoulder joint. Ensure they are placed with
a safe clearance to the C-arm. Do not leave the patient’s hands hanging outside the edges or under the
Base Board.
Kyphotic Patient
To accommodate Kyphosis of the spine you will need to place pillows under the patient’s torso to elevate. Adjust the Adjustable Assembly height and Face Pad Support angle if necessary. For extreme cases you may need to remove Head support completely and use only pillows.
Removing the Base Board
Ensure the product is removed from the table prior to moving the patient.
To remove the Base Board following a procedure, simply slide it out from underneath the mattress prior to moving the patient.
Disassembly
To remove, hold the Adjustable Assembly and tilt away from the arm board, before lifting out of the recess. The Face Pad can be removed from the Face Pad Support by removing the tips from the cavities. We recommend that the components from the Adjustable Assembly are not removed.
Prone Support Storage
We recommend detaching the Adjustable Assembly and Face Pad Support from the Base Board following a procedure for ease of storage.